Contents

Special Offer

From the Service Desk

Sensitivity & Late Eluting Compounds

Win Cinema Tickets

Alive and Kicking in Cheshire

Contacts & Comment

Click here if you have an application enquiry

We are always happy to get your feedback.

For a copy of our latest catalogue, click here.

Send to a Friend
Would your colleagues like a copy?
Previous Issues

July

August

September

  
From the Service Desk

 

In the first three issues of Peak Performance we looked at:

  • eliminating tailing on peaks by correct cutting of columns

  • aiding reproducibility through the use of a quartz wool plug in the liner

  • the relationship between flow rate and resolution.

This month I am exploring the importance of getting your sample into the column in the narrowest band possible and why decreasing the split ratio or injecting more sample may not be the answer.


Split ratio is the ratio of the carrier gas flow in the column and out of the split vent, with values typically ranging from 1:100 to 1:1000. A higher split ratio (e.g. 1:100 or 1%) introduces less sample into the column than a lower value (1:50 or 2%) but the actual amount of each sample compound entering the column varies and depends on many injector and carrier gas variables. The most important concern should be to obtain a consistent and reproducible split by calculating, measuring and setting the split ratio in the same manner for any given analysis.


Two things to note:

  • The split ratio will increase at higher oven temperatures. This is because gases become viscous at higher temperatures and so, for a given inlet pressure, the flow will decrease. A lower column flow means more split vent flow and, hence, a higher split ratio.

  • If there is a faulty liner seal then vapour can come back into the liner and circulate around the injector. This will cause an excessive tailing solvent peak and is symptomatic of a poor inlet liner seal.

Reducing the split ratio between 200:1 and 25:1 will have a linear effect on sample concentration. However, below 25:1 a linear increase in sample concentration does not result. For instance, going from 20:1 to 10:1 will not double the amount of sample on the column. As you can see on the graph, a ratio of 12:1 is more like 20:1. So if you want to accurately change the concentration of the sample on column keep the split ratio above 25:1.


In splitless mode, increasing the volume of the injection increases the amount of sample on the column. But once again, the increase is not linear. If you double the injection volume you will not double the amount of sample on column.

The main reason why this happens is that the liner cannot hold the entire volume of vaporised sample. Remember, the liner does not have a lid. So injecting more sample will just make more sample flow out the top of the liner. This always happens, even with small injection volumes, but the loss of sample becomes exaggerated when large volumes are used.

Not all solvents will expand to the same volume under a particular set of conditions. The volume of the expanded gas from a liquid of known volume and density can be calculated from the ideal gas law. It is important to remember that the volume of gas will increase as the molar mass decreases. This is one of the reasons why water is a very difficult solvent as it expands to 3 or 4 times the volume compared with methylene chloride.

Put simply, if the volume of the liner is not sufficient to handle the expanded gas volume, liner overload or Flashback will occur. Flashback can cause severe problems especially in quantitation. If semi volatile solutes enter the inlet lines, this can cause ongoing system contamination in the form of ghost peaks. A really bad case can result in the need to rip out the plumbing and replace the lines.

Flashback, however, can also be used to your advantage. If the injection system has become contaminated, water can be injected into the liner at high temperature and pressure deliberately to cause flashback and flush out contamination through steam distillation.

As you can see, it is not always possible to increase sensitivity by decreasing the split ratio (the peaks may get broader, costing you resolution, and making small peaks hard to see above the baseline), or by injecting more (because you may get flashback). If you need more sensitivity, are already using splitless injection or have the split ratio as low as you go and are injecting as much as you can, then the AirSharp may be just the answer.

If you have further questions, please send me an email or phone me on 01908 568844

Happy Chromatographing!


Anthony Addinall

[ click here to unsubscribe from this list ]